This is an HTML version of an attachment to the Official Information request 'SDS for Oral‑Fluid Devices & DNA‑Use Clarification'.


IR-01-26-21350 
9 July 2026 
Nigel Gray 
[FYI request #34578 email] 
Dear Nigel 
Request for information 
I refer to your Official Information Act 1982 (OIA) request of 14 June 2026 in which you 
asked for information regarding DrugWipe 3S Procurement and independent laboratory 
testing. I have addressed each part of your request below.  
On 11 June 2026, the Office of the Control er and Auditor-General (OAG) 
responded to my OIA request of 4 June 2026. In that response, the OAG stated: 
“The first, second and fifth categories of information you have requested wil  be 
held by New Zealand Police.” 
The categories referred to were: 
Al  documents, reports, and data relating to the independent laboratory testing 
conducted on DrugWipe 3S and other tenderers’ products. 
Al  tender evaluation documents relating to the DrugWipe 3S procurement. 
Any documents describing or summarising the testing process undertaken by 
Police or their contractors. 
However, in previous OIA responses, New Zealand Police have stated or implied 
that Police do not hold: 
independent laboratory testing reports 
validation or accuracy data 
testing methodologies 
pass/fail criteria 
scientific evaluation material for DrugWipe 3S 
Given the direct contradiction between the OAG’s statement and Police’s earlier 
responses, clarification is required. 
I therefore request the following under the OIA: 
1. Confirmation of whether New Zealand Police holds the information described
in categories 1, 2, and 5 above.
For clarity, this includes:
laboratory reports 
Police National Headquarters 
180 Molesworth Street. PO Box 3017, Wellington 6140, New Zealand. 
Telephone: 04 474 9499. Fax: 04 498 7400. www.police.govt.nz 



validation or accuracy data 
testing methodologies 
pass/fail criteria 
correspondence with laboratories or testing providers 
technical assessments 
scoring matrices 
evaluation panel notes 
conflict-of-interest declarations 
risk assessments 
internal or external summaries of the testing process 
2. If Police do hold this information, please provide it.
For tender evaluation documents, I refer you to pages 17-18 of the ‘Oral Fluid Testing 
Equipment Request for Proposal’ document which outlines the evaluation criteria and 
scoring matrix used during procurement. This document is publicly available; the link has 
been previously provided to you but has been reattached to this OIA for your reference 
here: https://www.police.govt.nz/about-us/publication/proactive-release-papers-relating-
implementation-roadside-drug-driving-testing  
For testing documents regarding the DrugWipe 3S Device, please also find included with 
my response the ‘IFC Expert Report’, and the ‘abf diagnostics Evaluation Report’.  
These reports contain technical scientific data that is generated and assessed in 
accordance with AS/NZS 4760:2019, the Standard that sets the performance and 
verification criteria for oral-fluid testing devices. To ensure the information is understood 
within the framework it was designed for, the response notes this context, and only the 
detailed dilution/spiking tables, containing non-public laboratory methodology, have been 
withheld under the relevant provisions of the OIA.  
As such, some of the information has been withheld pursuant to the fol owing sections of 
the OIA:  

9(2)(b)(i ), as the withholding of the information is necessary to protect the
commercial position of the person who supplied or who is the subject of the
information, where release would be likely to unreasonably prejudice that
commercial position (the withheld material comprises detailed, non-public
laboratory methods).

9(2)(ba)(i), as the withholding of the information is necessary to protect
information which is subject to an obligation of confidence or which any person
has been or could be compelled to provide under the authority of any enactment,
where the making available of the information would be likely to prejudice the
supply of similar information, or information from the same source, and it is in the
public interest that such information should continue to be supplied.

9(2)(k), as the withholding of the information is necessary to prevent the
disclosure or use of official information for improper gain or improper advantage
(the withheld material includes precise bench-level preparation recipes that could
be used to probe or game screening thresholds).




The fol owing certification was also provided to NZ Police for the DrugWipe 3 S: 

Compliance with EN ISO 10993 Biological evaluation of medical devices,
confirming there has no sensitizing potential.

Compliance with ISO 14971 Medical devices – Application of risk management
and ISO 10993 Biological evaluation of medical devices and confirmation of there
being no detection of harmful substances.
4. Any internal Police communications discussing the OAG’s statement that
Police hold this material.
Purpose of this request 
This request is necessary to reconcile: 
the OAG’s explicit statement that Police hold the independent laboratory testing 
and tender evaluation material, and 
Police’s earlier statements indicating that such material is not held. 
Police does not hold any communications discussing the Office of the Control er and 
Auditor General’s statements, including their statement to you that Police hold the 
independent laboratory testing and tender evaluation material. Their statement to you is 
referring you to Police as the appropriate agency to address your request. Therefore, 
your request is refused under section 18(e) of the OIA as the document alleged to contain 
the information requested does not exist. 
For clarity, Police has provided you publicly available information before regarding the 
DrugWipe 3 S device. In our letter to you dated 1 May 2026 (IR-01-26-11763), we 
confirmed that Police holds operational and procurement-related scientific validation and 
accuracy assessments of the DrugWipe 3 S. However, your request at the time was 
specifically for assessments in relation to legislative vetting or Bill of Rights Act (BORA) 
analysis which we had confirmed as not held by Police. 
I trust this information is satisfactory in answering your request. If you are not satisfied 
with the way your request has been responded to, you have the right under section 28(3) 
of the OIA to ask the Ombudsman to review our decisions. Information on how to do this 
is available online at www.ombudsman.parliament.nz. 
Yours sincerely 
Inspector Peter McKennie 
Acting Director: Road Policing 
New Zealand Police 


 
 
 
 
 
 
*** FINAL REPORT *** 
 
6 August 2025 
 
Phil Wheeler 
Procurement Lead 
Impaired Driving Programme 
National Road Policing Centre 
Police National Headquarters  
180 Molesworth St 
Wellington, New Zealand 
 
Re:  

 
Assessment of the Pathtech Roadside Drug Testing Device 
NZ Police: 
 
 
RFP TN/25/18 
IFC Expert Report No.:  
25-0105 
 
Dear Mr Wheeler,  
 
You  and  your  organisation  have  retained  Independent  Forensic  Consulting  represented  by  Dr 
Michael Robertson as consultants in toxicology and analytical chemistry in the above captioned 
matter.  
 
You have requested that I review the results of the Pathtech Securetec Drugwipe® 3 S drug device 
verification  and  comment  on  whether  the  results  of  the  on-site  testing  device  conform  to  the 
requirements  of  Appendix  C  of  the  current  oral  fluid  standard:  AS/NZS  4760  -  Procedure  for 
specimen collection and the detection and quantitation of drugs in oral fluid (the Standard).  
 
 
  
 
 
 
Expert Report No. 25-0105 
Page 1 of 7 




METHODS / PROCEDURE 
 
Device verified - Pathtech: Securetec Drugwipe 3 S. Batch Y1647.25092, Expiry 2026-02. 
 
The device was assessed against the screening cut-offs for on-site immunoassay screening test 
listed in Table A1; Appendix A of AS/NZS 4760:2019 and are as follows: 
 
 
Stock solutions of the required calibrators (methamphetamine, cocaine or benzoylecgonine and 
THC) were prepared immediately prior to performing the verification.  
 
Blank human oral fluid specimens were spiked with the appropriate volume of the stock solutions, 
one spiked at +50% of the screening cut-off concentrations and the other at -50% of the screening 
cut-off concentrations.  
 
A separate set of stock solutions of MDMA were also prepared immediately prior to performing 
the verification.  
 
The concentrations of the MDMA stock solutions prepared were dependent on the cross reactivity 
of MDMA on the methamphetamine strip of the respective device.  
 
In  order  to  test  to  section  C3  of  Appendix  C  of  AS/NZS  4760:2019,  device  verification  was 
assessed by testing a minimum of twenty (20) spiked samples. Ten (10) kits with oral fluid samples 
spiked at -50% of the screening cut-offs, and ten (10) kits with oral fluid samples spiked at +50% 
of the screening cut-offs.  
 
All testing was performed in accordance with the provided manufacturer’s instructions. 
 
Section C3 of Appendix C states that if a total of twenty kits are tested, no more than two failures 
in total (either false ‘not-negative’ or false ‘negatives’) are permitted for each drug class tested.  
 
A  separate  verification  was  performed  for  MDMA  due  to  its  cross  reactivity  on  the 
methamphetamine strip.  
 
Prior to testing, all components were brought to room temperature prior to verification.  
 
Expert Report No. 25-0105 
Page 3 of 7 


Fresh oral fluid was collected from drug free human donors. Specimens were pooled, centrifuged 
and frozen until required for evaluation. Centrifugation was required to remove bubbles etc. to 
allow for a more accurate measurement of oral fluid volume, an important aspect of verification.  
 
To ensure the correct concentration of drugs were spiked into the samples, aliquots of the blank 
oral fluid, -50% spiked oral fluid and +50% spiked oral fluid were also assessed by LCMS.  
 
The  Pathtech:  Securetec  Drugwipe®  3  S  device  has  the  following  drug  test  strips  and  cut-off 
concentrations, all of which conform to the Standard: 
 
Aliquots  (3  x  10μL)  of  the  appropriate  spiked  oral  fluid  specimen  were  added  to  the  sample 
collector pads before being placed back onto the test device. The test device was held upright, a 
thumb  placed  over  the  word  PRESS  before  pressing  down  hard  once,  breaking  the  internal 
ampoule. The device was held upright for a further 10 seconds before being placed on the bench 
surface (Instructions for Use, S 304 G.91). 
 
A control line must be observed for the test to be valid. Any visible line or part thereof on the test 
strip indicates a positive or non-negative result. No visible line indicates a negative result. All 
results were read at 5 minutes after the test was started.  
 
The temperature during verification was recorded by a calibrated data logger and was 23.2°C. 
 
 
 

 
Expert Report No. 25-0105 
Page 4 of 7 



RESULTS 
 
1.  All compounds and cutoff concentrations comply with AS/NZS 4760:2019. 
 
2.  The results for the evaluation of the Pathtech: Securetec Drugwipe® 3 S device are shown in 
the tables below. 
 
 
3.  Control lines were present for all observed tests.  
 
4.  There were two false positives for methamphetamine and one false positive each for THC and 
MDMA, however this is within the required criteria of the Standard.  
 
5.  Concentrations of the spiked oral fluid samples were confirmed by LCMS and were within 
acceptable ranges for all drug groups. 
 
 
CONCLUSION 
 
The  Pathtech:  Securetec  Drugwipe®  3  S  device  was  verified  and  complied  with  the  criteria 
outlined in Appendix C3 of AS/NZS 4760:2019 for methamphetamine, cocaine, THC and MDMA.  
 
The Pathtech: Securetec Drugwipe® 3 S device has been tested and does comply with the Standard 
and  therefore  does  meet  the  required  cut-off  thresholds  defined  in  Table  A1  of  the  joint 
AS/NZS4760:2019  Standard  and  it’s  cut  off  thresholds  and  false  positive/negative  results  as 
required by pre-condition 5 of NZ Polices OFTE RFP. 
 
 
Sincerely, 
 
 
 
 
 
Dr Michael Robertson 
Pharmacologist, Forensic Toxicologist, Analytical Chemist 
 
 

 
Expert Report No. 25-0105 
Page 5 of 7 

Qualifications, experience and expertise in the field of drug and alcohol policies, 
procedures and testing 
I am a pharmacologist and forensic toxicologist at Independent Forensic Consulting (IFC).  After 
completing  my  undergraduate  degree  majoring  in  pharmacology  and  toxicology,  I  earned  my 
Ph.D. in medicine studying at the Victorian Institute of Forensic Medicine (Monash University), 
specialising  in  toxicology.    I  subsequently  completed  my  post-doctoral  fellowship  in  forensic 
toxicology at National Medical Services (NMS) in Pennsylvania (USA).  
As a pharmacologist and forensic toxicologist with more than 30 years professional experience, I 
have  studied  the  affects  and  effects  of  drugs  and  poisons  on  humans  and  animals  including 
mechanism of action; desirable effects and non-desirable or adverse effects.  I have also performed 
research in the field of pharmacology and toxicology including methods of drug analysis; affects 
of drugs on humans; evaluation of the safety and efficacy of drugs on humans.  I have assigned, 
supervised, performed, certified and interpreted hundreds of toxicological analyses for a range of 
clients  including  pathology  and  forensic  laboratories,  pharmaceutical  companies,  regulatory 
agencies and legal professionals.   
I  have  prepared  expert  reports  and  testified  throughout  Australia  and  overseas  in  local,  state, 
federal  and  military  courts  for  prosecution,  defence  and  plaintiff  lawyers  relating  to  forensic 
toxicology including matters involving: 
• Criminal Law (including drug facilitated sexual assault; DUI and DUID; cause or contribution
to death; human performance and behaviour; poisonings; drug possession and matters of legal
status of drugs including the evaluation of Synthetic Cannabis and Synthetic Stimulants e.g.
‘Bath Salts’ and issues associated with analogue provisions);
• Family Law (including the interpretation of urine, drug and hair follicle testing; Liver Function
Tests (LFT), Carbohydrate Deficient Transferrin (CDT) and Ethyl Glucuronide (EtG) testing
and result interpretation);
• Personal Injury, Medical Negligence, CTP, Insurance claims;
• Workplace Disputes (including the interpretation of urine, oral fluid and/or hair test results and
compliance with acceptable drug testing procedures);
• Workplace  Accidents  (involving  the  cause  and  contribution  of  drugs  and  alcohol  to  an
incident); together with matters including exposure to chemicals and other poisons;
• Doping  disputes  and  investigations  including  professional  athletes  (ASADA;  WADA)  and
Steward enquiries involving the racing industry (jockeys, riders, racehorses, harness racing,
greyhounds etc.);
• Miscellaneous civil matters and disputes involving drugs and chemicals.
I have lectured at numerous Universities on topics such as pharmacology and forensic toxicology.  
I have authored a number of peer-review papers, and I routinely attend national and international 
scientific conferences.  I currently hold membership of The International Association of Forensic 
Expert Report No. 25-0105 
Page 6 of 7 

Toxicologists (TIAFT), the Society of Hair Testing (SoHT), the Society of Forensic Toxicologists 
(SOFT), The Australian and New Zealand Forensic Science Society (ANZFSS) and the Forensic 
and  Clinical  Toxicology  Association  (FACTA)  and  am  bound  by  the  Code  of  Ethics  of  these 
Societies.  I have been a member of the SOFT Drugs and Driving Committee and SOFT Drug 
Facilitated Sexual Assault Committee and served as an invited reviewer on the SOFT Scientific 
Advisory panel.  I currently serve as an invited reviewer for the international journals, Forensic 
Science International and Forensic Science, Medicine, and Pathology.   
 
With respect to workplace drug testing, I am a consulting toxicologist to a number of national drug 
testing organisations and an expert in Australian Standard compliant drug testing for both urine 
and oral fluid (AS/NZS 4308:2023 and AS/NZS 4760-2019 respectively).  I am the immediate 
past Chair of the Standards Australia oral fluid committee reviewing and updating AS/NZS 4760 
– Procedure for specimen collection and the detection and quantitation of drugs in oral fluid.  I am 
the  immediate  past  Chair  of  the  Standards  Australia  urine  committee  reviewing  and  updating 
AS/NZS 4308 – Procedure for specimen collection and the detection and quantitation of drugs in 
urine.  I  am  the  current  Chair  of  Standard  Australia’s  technical  committee  (CH-039)  that  is 
responsible  for  the  ‘Detection  and  Analysis  of  Drugs  in  Specimens  of  Human  Origin’. I  am  a 
contributing member of the FACTA technical committee on Hair Drug Testing Guidelines. 
 
I have been involved in and testified in a number of workplace disputes before Fair Work Australia 
and other tribunals including: Endeavour Energy and CEPU, ASU and APESMA; The Maritime 
Union of Australia v DP World Brisbane Pty Ltd and Ors; and CFMEU v PKCT; among others. 
 
As a qualified trainer in the workplace collection and testing of breath, urine and oral fluids in 
compliance  with  AS  3547:2019,  AS/NZS  4308:2008;  AS/NZS  4308:2023;  AS  4760-2006  and 
AS/NZS 4760-2019 respectively, I perform drug and alcohol training throughout Australia for a 
number  of  clients  to  ensure  compliance  with  the  Australian  Standards.    This  training  includes 
ensuring compliant collection, on-site testing, storage, and transportation to the laboratory. Policy 
review  and  recommendations  to  ensure  the  respective  drug-testing  program  in  place  is  both 
compliant to the respective Australian Standard and is able to achieve the desired outcome or intent 
of the client.  In the last 10 years I have performed more than 200 training sessions across more 
that 150 organisations.  
 
Expert Report No. 25-0105 
Page 7 of 7 


 abf diagnostics GmbH 
Gewerbepark 14 
D-85402 Kranzberg
Germany 
T.: +49/8166-9986-130 
F.: +49/8166-9986-140 
E.: [email address] 
Kranzberg, 14th of March 2025 
Evaluation Report - Securetec DrugWipe 3 S (S303G.91) 
Introduction 
DrugWipe  3 S test kits were evaluated for specificity and sensitivity in compliance around 
nominated cut-offs for each of the three analyte groups in accordance to Appendix C3 auf AS/NZS 
4760:2019 with the following protocol.  
Methods: 
Drug solutions were prepared by spiking pooled saliva of 10 volunteers with the analytes. The 
final dilutions provide a test sample with a concentration 50 % above (high drug “spiked” solution) 
or 50 % below (lower drug “spiked” solution) the specified cut-off.  
s 9(2)(b)(ii) OIA, s 9(2)(k) OIA, s 9(2)(ba)(i) OIA
abf diagnostics – Drugwipe Evaluation Report 
Page 1 of 4 

s 9(2)(b)(ii) OIA, s 9(2)(ba)(i) OIA, s 9(2)(k) OIA 
 
 
 
 
 
 
 
 
    
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
    
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
    
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
    
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
    
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
    
 
 
 
 
 
 
 
 
 
abf diagnostics – Drugwipe Evaluation Report 
Page 2 of 4 
 

Specificity testing: 
10 µL of each lower drug “spiked” solution was pipetted onto the respective wiping fleece of the 
DrugWipe® sample collectors and tests were performed according to the enclosed instructions 
for use. The tests were evaluated visually as described in the instructions for use. 
An appearing red control line and no red test line shows a negative result. 
Sensitivity testing: 
10 µL of each high drug “spiked” solution was pipetted onto the respective wiping fleece of the 
DrugWipe® sample collectors and tests were performed according to the enclosed instructions 
for use. The tests were evaluated visually as described in the instructions for use. 
An appearing red control line and a red test line shows a positive result. 
Results: 
Specificity at – 50% cut-off 
Analyte 
Concentration 
Negative results 
% Correct results 
Δ9-THC 
7 ng/ml 
27/30 
90% 
D-
25 ng/ml 
27/30 
90% 
Methamphetamine/ 
D.L-MDMA
Cocaine 
25 ng/ml 
29/30 
96% 
Sensitivity at + 50% cut-off 
Analyte 
Concentration 
Positive results 
% Correct results 
Δ9-THC 
22 ng/ml 
28/30 
93% 
D-
75 ng/ml 
30/30 
100% 
Methamphetamine/ 
D.L-MDMA
Cocaine 
75 ng/ml 
29/30 
96% 
abf diagnostics – Drugwipe Evaluation Report 
Page 3 of 4 


Summary: 
 
DrugWipe 3 S (S304G.91) test kits fulfil ed the criteria of Appendix C3 of AS/NZS 4760:2019 with 
not more than a total of 10% false positive results in specificity testing at – 50% of the respective 
cut-offs and not more than a total of 10% false negative results in sensitivity testing at + 50% of 
the respective cut-offs for THC, Methamphetamine/MDMA and Cocaine. 
 
 
 
Kranzberg, 14th of March 2025 
 
 
 
 
 
 
………………………………. 
Dr. Franz Aberl 
abf diagnostics GmbH 
 
abf diagnostics – Drugwipe Evaluation Report 
Page 4 of 4